In its simplest form, spirometry is the measurement of how much air a person can breathe out in one forced breath, known as forced expiratory vital capacity (FVC). Different parameters can be measured from the flows that occur during the FVC maneuver. Healthcare professionals interpret volume and flow data alongside graphs showing volume/time and flow/volume to assess lung health.
Vitalograph has investigated many different precision technologies for measuring flows and volumes. Today, the preferred technology used in Vitalograph diagnostic spirometers for flow measurement is the Fleisch Pneumotachograph, and for volume measurement is the Rolling Seal. We use Lilly-type screen pneumotachographs in our lab spirometers and stator-rotor flow measuring technology in our monitoring and screening devices.
Fleisch pneumotachograph flow measuring technology is the most accurate, precise and reliable flow measuring technology. The linear signal is simple to process and the spirometry device is extremely robust as it has no moving parts. Even with a Bacterial Viral Filter (BVF), essential for effective hygiene between the patient and the measuring system, the system flow impedance is low even at high flow rates and well within the requirements of the current spirometer standard ISO26782:2009. Fleisch flowmeters may be heated or unheated, but in office spirometry flowmeters are generally unheated for safety, speed of use and accuracy. Vitalograph spirometers use an unheated Fleisch Pneumotachograph.
A Fleisch Pneumotachograph consists of a bundle of small capillary tubes which create pseudo-laminar flow. The result of this is a linear output pressure signal created by increasing air flow. Two annular rings of pressure measuring points in the Fleisch are ported out to an extremely low range differential pressure transducer, operating in the linear range of the transducer. The differential pressure is sampled by the associated electronics at a frequency of at least 100Hz and 14 bits to allow the instantaneous and accurate calculation of flow rates and, by differentiation over time, the accumulated volume. The turndown ratio of such a device is a staggering 20:1, which allows accurate measurement of very low flows as well as being capable of measuring very high flows.
Small and lightweight, the simple digital signal produced by the spin of the rotor provides accurate results for monitoring and screening of respiratory parameters such as PEF, FEV1 and FEV6. On Vitalograph devices, a simple LCD screen provides immediate results and the device stores several hundred tests that can be downloaded to view trends in patient data. This spirometric measuring technology is often referred to as a ‘turbine’, although a true turbine has a curved blade and requires no stator.
View our range of monitors and screeners which use this technology.
In a stator-rotor system, a multi-vaned stator induces swirl on the airflow which makes a flat blade (the rotor) spin. The rotor interrupts a light source as it spins, and a detector measures the air flow by counting the revolutions. The rotor rotates at a speed that is approximately proportional to flow. Devices are calibrated at the time of production.
Stator-rotor and turbine based systems are not a suitable technology for measuring very low flows because of the effect of inertia on the spin of the rotor, making it unsuitable for full spirometry. Rather than risking inaccurate FVC readings, Vitalograph devices use a proven surrogate measure, FEV61 ,2, in our low cost monitoring and screening devices that use stator rotor flow measuring technology.
1 Bellia V, Sorino C, Catalano F, et al Validation of FEV6 in the elderly: correlates of performance and repeatability Thorax 2008;63:60-66.
2 M. Akpinar-Elci, K. B. Fedan, P. L. Enright FEV6 as a surrogate for FVC in detecting airways obstruction and restriction in the workplace European Respiratory Journal Feb 2006, 27 (2) 374-377.
Direct determination of volume has the advantage of being absolute and measured at ambient conditions during both inspiration and expiration. This high accuracy makes the rolling seal well suited to routine clinical testing, research and as an accuracy reference for other non-volumetric spirometers. Volumetric devices are also ideal for training where direct determination of volume provides a clear platform for demonstration and explanation of technique and results.
Launching soon! The Gold Standard, our new rolling seal spirometer, Talk to us to learn more.
A lightweight piston is mounted horizontally in a cylinder and a rod that rests on frictionless bearings supports the piston which is attached to the cylinder wall by a flexible plastic seal. When the test subject completes a spirometry blow, the piston moves by rolling the seal onto itself rather than sliding, in direct proportion to the volume of exhaled air. The large diameter piston only requires movement of a few inches to record a large volume change. A highly accurate digital signal is generated via a potentiometer to a PC for viewing and storing data. In order to ensure smooth, accurate flow profiles, Vitalograph have developed an exceptionally low resistance movement in the Gold Standard rolling seal mechanism.
Lilly screen pneumotachographs measure flow in a similar way to Fleisch technology, employing a fine wire mesh screen with low resistance to air flow. Flow is calculated by measuring the pressure differential across a screen or series of screens. This type of device is very accurate and stable. A screen pneumotachometer is slim, allowing space in the flowhead for other features e.g. demand valves and other gas inputs and outputs making it suitable for lab spirometry requiring additional gases for tests such as DLCO and nitrogen washout.
The screen used in a Lilly-type Pneumotachograph is very easy to clean and service. The heated screen provides stable measurement for protracted periods of testing, although the additional electronic components in the flowhead render it less suitable for office spirometry than Fleisch flowhead technology.
While this technology is very sensitive, Vitalograph does not use ultrasonic flow measuring technology as it is prone to errors before measurement with low flow profiling from ambient air movement and does not measure low flows accurately3. Accuracy depends on a developed flow profile that is not found in turbulent expired air, therefore flow rates must take into account the turbulent gas flow profile; in ultrasonic technology this is corrected by software algorithms based on Gaussian probability. Turbulent airflow is overcome in Fleisch, Screen and Stator-Rotor and turbine measuring technology by design elements that manage airflow i.e. the swirl induced by the stator in stator-rotor devices, laminar flow induced by capillaries in the Fleisch and the mesh in Lilly technology.
Hygiene in ultrasonic devices is dependent on expensive disposable flow tubes with additional filters required for infection control.
3 Perera, A., Friedrich, P., & Ledermüller, R. (2018). Novel hot-wire based spirometry is highly accurate at low flow rates, Current Directions in Biomedical Engineering, 4(1), 513-515.
In an ultrasonic flow meter, the distance that an ultrasonic pulse has to travel between transducers is known, as is the time taken for the pulse to move between transducers with no gas flow at known temperature, pressure and humidity. Changes in gas flow create measurable changes in the time taken for the ultrasonic pulses to travel between transducers and this difference creates an analog signal that can be measured to calculate gas flow rate through the spirometer.
The time taken for the ultrasonic pulse to move between transducers is affected by temperature, humidity and ambient flows which requires these factors to be measured and correction algorithms applied.
Hot wire measuring technology is used in lab spirometer systems as it is capable of measuring flow using gases other than room air by applying conversion factors and it can give very stable, accurate measurements. However the requirement for delicate electrical wires within the flowhead means that the technology is unsuited to portable applications and can be unreliable. Vitalograph does not use hot wire measuring technology due to its fragility and the added complexity of patient safety incurred by including electrical parts within the flowhead.
Gas flow past an electrically heated wire element causes a temperature decrease in proportion to the mass of gas and its flow. Due to this, the resistance of the wire also changes. The change in resistance of the wire becomes a measure of flow rate. The flow signal is integrated electronically or by software to obtain a volume measurement.
Vitalograph is a global leader in respiratory diagnostics. We produce high quality, accurate and reliable, respiratory medical devices & services for primary, secondary healthcare and occupational health. We deliver successful clinical trials for leading pharmaceutical companies, biotechs and research organisations.
© Vitalograph 2011 – 2021 | Privacy Policy | Cookies
Vitalograph Ltd
Vitalograph Business Park
Maids Moreton
BUCKINGHAM
MK18 1SW
England
T: +44(0) 1280 827110
Clinical Trials: clinical.trials@vitalograph.co.uk
Healthcare: sales@vitalograph.co.uk