Extended Warranty
Free 5 year warranty upon device registration
Diagnose, monitor and manage patients’ respiratory health in the most efficient and effective way possible.
Custom solutions and a personalized approach mean we can meet a wide range of respiratory clinical study needs.
The latest respiratory news including clinical research, articles on respiratory topics, patient information pieces, and episodes of our ever popular Exhale podcast.
Vitalograph products and services are chosen by medical professionals in over 113 countries for healthcare, occupational health, and clinical trials.
Ideal for monitoring PEF and FEV1 at home, work, in clinical trials, in primary or secondary care.
The Vitalograph asma-1™ is a lightweight home use electronic respiratory monitor that measures PEF and FEV1. It offers greater accuracy than mechanical peak flow meters and eliminates the need for paper record cards. Links easily to other devices by USB for reporting.
Measures FEV1 and PEF.
Displays % of personal best FEV1.
Electronic record – no need for record cards.
Set unique ‘personal best’ zones for test subjects.
Automatically stores best values in a test session.
Quality of blow indicator.
600 test memory.
Suitable for use with Bacterial Viral Filters (BVF™).
Compliant with: ATS/ERS 2019 Technical Standards and ISO 23747 / 26782 / 13485
Large, easy to read display.
Includes Vitalograph Reports to generate PDF reports to PC with:
Information presented in tables and graphs.
Customizable report headers.
Ability to add comments to the report.
Subject ID and name fields.
Age, height, gender, and weight fields optional.
One of our 4000 Series of respiratory monitors and screeners, the asma-1 comes with:
Pouch
Instructions for use
USB cable
2 x AAA Batteries
Plastic mouthpiece for single patient use
Vitalograph Reports Software
Vitalograph provides free certified training for all of its respiratory diagnostic solutions. Register here.
Click here for reimbursement codes for Vitalograph Devices in the USA.
Product Code: 40400
Respiratory Monitor asma-1 USB | |
Model | 4000 |
Dimensions | 109mm (length) x 63mm (width) x 42mm (height) |
Accuracy | Better than ± 3% (FEV1), ± 10% (PEF) |
Flow Detection Principal | Stator rotor |
Back Pressure | Less than 0.15kPa/L/second @ 14L/s |
Measurement Range | PEF: 25 – 840 L/min BTPS FEV1: 0 – 9.99 L BTPS |
Maximum Test Duration | 1 second |
Quality Check | Time to Peak Flow >120ms or a cough detected in the first second |
Power Supply | 3V (2 x 1.5V AAA batteries) |
Battery Life | 3 months of use, 3 tests per day |
Performance Standards & Guidelines | ATS/ERS 2019 Technical Standards, ISO 23747:2015, ISO 26782:2009 |
Communications | USB2.0/3.0 |
PC Requirements | Windows® 7, Windows® 8 or Windows® 10 |
Auto Power Down Time | 2 minutes |
asma-1 | asma-1 Child | asma-1 USB/BT | copd-6 | copd-6 USB/BT | lung monitor | lung monitor USB/BT | lung monitor BT Smart | lung age | |
---|---|---|---|---|---|---|---|---|---|
PEF | ✓ | ✓ | ✓ | ✓ | |||||
FEV1 | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
FEV6 | ✓ | ✓ | ✓ | ✓ | ✓ | ||||
FEV1 Ratio | ✓ | ✓ | ✓ | ✓ | ✓ | ||||
FEF25-75 | ✓ | ✓ | |||||||
FEV0.5/0.75 | ✓ | ✓ | |||||||
Lung Age | ✓ | ✓ | ✓ | ||||||
Memory | 600 | 600 | 600 | 0 | 0 | 200 | 200 | 600 | 0 |
Personal best | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ | ✓ |
GOLD | ✓ | ✓ | |||||||
PDF Reports | ✓ | ✓ | ✓ | ||||||
Import Data to SpV | ✓ |
Can I use a Bacterial Viral Filter (BVF) with the asma-1 USB?
AddA BVF is recommended to use if the asma-1 USB is intended for multi-subject use. The plastic mouthpiece supplied with the device can be used where the device is intended for single-subject use. For subject groups with excessive sputum production, a BVF can be used to protect the device flowhead from exhaled particulate matter where this may be a concern.
What does the quality of blow indicator detect?
AddThe (!) appears after a blow if; Vext (extrapolated volume) is > 5% or 100mL of FEV6 or a cough is detected in the 1st second.
How do you clean the asma-1 USB?
AddIf a new BVF is used for every test subject and the asma-1 USB is stored in a clean, dust free environment, then the device should require only surface cleaning using a 70% isopropyl alcohol wipe. If you suspect that the device flowhead has become contaminated, refer to the 4000 Series Cleaning Instructions.
Can the asma-1 connect with my phone/tablet?
AddThe asma-1 USB connects to your PC using the USB cable supplied.
To connect to a supported mobile App, you require the lung monitor BT Smart. For more details talk to us.
All Vitalograph spirometers come with a host of free benefits that are activated when you register your device.
Free 5 year warranty upon device registration
New features and improvements, automatically
Online device training - unlimited, free, & forever
Expert support for Vitalograph users
As experts in Respiratory Diagnostics, we are committed to providing our community with the latest in educational material.